GENERAL |
|
|
|
|
Analysis and Verification of the Relationship between the Maximum Mass Sensitivity of Quartz Crystal Microbalance and Electrode Parameters |
Wei Pan, Xian-He Huang**, Qiao Chen, Zhi-Chao Fan, Yuan Xu |
School of Automation Engineering, University of Electronic Science and Technology of China, Chengdu 611731
|
|
Cite this article: |
Wei Pan, Xian-He Huang, Qiao Chen et al 2019 Chin. Phys. Lett. 36 070701 |
|
|
Abstract We analyze the effect of electrode diameter and thickness on the mass sensitivity. Through the theoretical approximate calculation, we find that the mass sensitivity does not change monotonically with electrode diameter and there is a maximum point. The optimum electrode diameter corresponding to the maximum mass sensitivity varies with the electrode thickness. For a particular electrode diameter, a quartz crystal microbalance (QCM) with thick electrode has a higher mass sensitivity. A proper plating experiment using 35 QCMs with different electrode diameters and thicknesses verifies this finding. The present study further reveals how electrode size affects mass sensitivity and is helpful for QCM design.
|
|
Received: 16 March 2019
Published: 20 June 2019
|
|
PACS: |
07.07.Mp
|
(Transducers)
|
|
07.75.+h
|
(Mass spectrometers)
|
|
43.25.Zx
|
(Measurement methods and instrumentation for nonlinear acoustics)
|
|
43.35.Ns
|
(Acoustical properties of thin films)
|
|
|
Fund: Supported by the National Natural Science Foundation of China under Grant No 61871098. |
|
|
[1] | Sauerbrey G 1959 Physik 155 206 | [2] | Yao Y, Chen X D, Li X Y, Chen X P and Li N 2014 Sens. Actuators B 191 779 | [3] | Sun Y G and Wang H H 2007 Adv. Mater. 19 2818 | [4] | Pei Z F, Ma X F, Ding P F, Zhang W M, Luo Z Y and Li G 2010 Sensors 10 8275 | [5] | Li X Y, Chen X D, Yao Y, Li N and Chen X P 2014 Sens. Actuators B 196 183 | [6] | Kimura M, Liu Y, Sakai R, Sato S, Hirai T, Fukawa T and Mihara T 2011 Sens. Mater. 23 359 | [7] | Wang X F, Ding B, Sun M, Yu J Y and Sun G 2010 Sens. Actuators B 144 11 | [8] | Shi S, Reisberg S, Anquetin G, Noël V, Pham M C and Piro B 2015 Biosens. Bioelectron. 72 205 | [9] | Escobar-Teran F, Arnau A, Garcia J V, Jiménez Y, Perrot H and Sel O 2016 Electrochem. Commun. 70 73 | [10] | Yadav S K, Agrawal B, Chandra P and Goyal R N 2014 Biosens. Bioelectron. 55 337 | [11] | Lai G S, Zhang H, Tamanna T and Yu A 2014 Anal. Chem. 86 1789 | [12] | Miguel F H C, Benedetti T M, Torresi R M and de Torresi S I C 2014 Electrochem. Commun. 48 164 | [13] | Levi M D, Shpigel N, Sigalov S, Dargel V, Daikhin L and Aurbach D 2017 Electrochim. Acta 232 271 | [14] | Lu Y C, Song S Y, Hou C X, Pang S, Li X M, Wu X W, Shao C, Pei Y X and Pei Z C 2018 Chin. Chem. Lett. 29 65 | [15] | Wu Y F, Xue P, Kang Y J and Hui K M 2013 Anal. Chem. 85 3166 | [16] | Zhang S L, Bai H H, Pi J, Yang P H and Cai J Y 2015 Anal. Chem. 87 4797 | [17] | Jia X E, Zhang Z W, Tan L, Zhang Y Y, Xie Q J, He Z M and Yao S Z 2006 Chin. Chem. Lett. 17 509 | [18] | Kovář D, Farka Z k and Skládal P 2014 Anal. Chem. 86 8680 | [19] | Lee J, Atmeh M and Berman D 2017 Carbon 120 11 | [20] | Chen D Q, Sun X Y, Zhang K H, Fan G K, Wang Y, Li G and Hu R F 2017 Sensors 17 1681 | [21] | Hu R F, Zhang K H, Fan G K, Luo Z Y and Li G 2015 Meas. Sci. Technol. 26 055102 | [22] | Kurosawa S, Park J W, Aizawa H, Wakida S I, Tao H and Ishihara K 2006 Biosens. Bioelectron. 22 473 | [23] | Wei X R, Qu X L, Ding L, Hu J T and Jiang W 2016 Environ. Pollut. 219 1 | [24] | Wang J, Yang J S and Li J Y 2007 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 54 687 | [25] | Wang J, Shen L J and Yang J S 2008 Ultrasonics 48 150 | [26] | Shockley W, Curran D R and Koneval D J 1967 J. Acoust. Soc. Am. 41 981 | [27] | Huang X H, Pan W, Hu J G and Bai Q S 2018 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 65 1888 | [28] | Mecea V M 2006 Sens. Actuators A 128 270 | [29] | Josse F, Lee Y, Martin S J and Cernosek R W 1998 Anal. Chem. 70 237 | [30] | Gao J Y, Huang X H and Wang Y 2013 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 60 2031 | [31] | Hillier A C and Ward M D 1992 Anal. Chem. 64 2539 | [32] | Ward M D and Delawski E J 1991 Anal. Chem. 63 886 | [33] | Huang X H, Bai Q S, Hu J G and Hou D 2017 Sensors 17 1785 | [34] | Huang X H, Bai Q S, Zhou Q and Hu J G 2017 Sensors 17 1476 |
|
|