NUCLEAR PHYSICS |
|
|
|
|
Measured and Simulated Performance of a Four-Segmented Clover Detector |
ZHANG Ning-Tao1**,LEI Xiang-Guo1,ZHANG Yu-Hu1,ZHOU Xiao-Hong1,LIU Min-Liang1,GUO Ying-Xiang1,FANG Yong-De1,LI Shi-Cheng1,2,ZHOU Hou-Bing1,2,DING Bing1,2,WANG Hai-Xia1,2,WU Xiao-Guang3,HE Chuang-Ye3,ZHENG Yun3 |
1Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000
2Graduate School of Chinese Academy of Sciences, Beijing 100049
3China Institute of Atomic Energy, Beijing 102413 |
|
Cite this article: |
ZHANG Ning-Tao, LEI Xiang-Guo, ZHANG Yu-Hu et al 2012 Chin. Phys. Lett. 29 042901 |
|
|
Abstract Gamma rays with energies between 80 keV and 1500 keV produced by 133Ba and 152Eu standard sources are measured with the clover detector. Relative efficiencies and add-back factors are determined in both crystals and full clover modes. The add-back characteristics have been calculated using Monte-Carlo simulation code Geant4. The simulation and measurements agree very well.
|
|
Received: 06 January 2012
Published: 04 April 2012
|
|
|
|
|
|
[1] Beausang C W and Simpson J 1996 J. Phys. G 22 527[2] Svensson C E et al 2003 Nucl. Instrum. Methods B 204 660[3] Lee I Y, Deleplanque M A and Vetter K 2003 Rep. Prog. Phys. 66 1095[4] Eberth J and Simpson J 2008 Prog. Part. Nucl. Phys. 60 283[5] Gerl J 2000 Nucl. Instrum. Methods A 442 238[6] Duchêne G et al 1999 Nucl. Instrum. Methods A 432 90[7] France G de 2004 Eur. Phys. J. A 20 59[8] GANIL http://pro.ganil-spiral2.eu/laboratory/detectors/exogam[9] Ball G C et al 2007 Nucl. Phys. A 787 118[10] Scraggs H C et al 2005 Nucl. Instrum. Methods A 543 431[11] Lipoglavšek M et al 2006 Nucl. Instrum. Methods A 557 523[12] Muralithar S et al 2010 Nucl. Instrum. Methods A 622 281[13] Agostinelli S et al 2003 Nucl. Instrum. Methods A 506 250[14] Schumaker M A et al 2007 Nucl. Instrum. Methods A 570 437 |
|
Viewed |
|
|
|
Full text
|
|
|
|
|
Abstract
|
|
|
|
|