Abstract :The NIM5 fountain clock is the second fountain clock built at NIM (National Institute of Metrology, China), and has been operating stably and sub-continually since 2008. The fountain operates with a simple one-stage optical molasses to collect cold atoms, which reduces the collisional frequency shift dramatically. The fractional frequency uncertainty is estimated to be 2×10?15 . The typical frequency instability of 2.5×10?14 is obtained at 10 s. Comparisons with other fountain frequency standards worldwide demonstrate agreement within the stated uncertainties.
收稿日期: 2012-08-01
出版日期: 2013-03-04
:
06.30.Ft
(Time and frequency)
43.58.Hp
(Tuning forks, frequency standards; frequency measuring and recording instruments; time standards and chronographs)
95.55.Sh
(Auxiliary and recording instruments; clocks and frequency standards)
引用本文:
. [J]. Chin. Phys. Lett., 2013, 30(1): 10601-010601.
LIU Nian-Feng, FANG Fang, CHEN Wei-Liang, LIN Ping-Wei, WANG Ping, LIU Kun, SUO Rui, LI Tian-Chu. Accuracy Evaluation of NIM5 Cesium Fountain Clock. Chin. Phys. Lett., 2013, 30(1): 10601-010601.
链接本文:
https://cpl.iphy.ac.cn/CN/10.1088/0256-307X/30/1/010601
或
https://cpl.iphy.ac.cn/CN/Y2013/V30/I1/10601
[1] Li T C, Li M S, Lin P W, Wang P, Chen W L, Liu N F, Lin Y G 2007 Chin. Phys. Lett. 24 1177 [2] Li T C, Li M S, Lin P W, Wang P, Chen W L, Liu N F and Lin Y G 2007 IEEE International Frequency Control Symposium, 2007 Joint with the 21st European Frequency and Time Forum p 55 [3] Tiesinga E, Verhaar B J, Stoof H T C and van Bragt D 1992 Phys. Rev. A 45 R2671 [4] Leo P J, Julienne P S, Mies F H and Williams C J 2001 Phys. Rev. Lett. 86 3743 [5] Douglas R J, Thomas C 1997 29th Annual Precise Time and Time Interval (PTTI) Meeting (California, USA 2–4 December 1997) p 85 [6] Szymaniec K, Cha?upczak W, Whibberley P B, Lea S N and Henderson D 2005 Metrologia 42 49 [7] Weyers S, Hubner U, Schroder R, Tamm C and Bauch A 2001 Metrologia 38 343 [8] Levi F, Calonico D, Lorini L and Godone A 2006 Metrologia 43 545 [9] Liu N F, Li T C, Lin P W, Wang P, Chen W L and Lin Y G 2010 Acta Metrologica Sin. 31 274 [10] Vanier J and Audoin C 1989 The Quantum Physics of Atomic Frequency Standards (Bristol: Hilger) [11] Wynands R and Weyers S 2005 Metrologia 42 S64 [12] Simon E, Laurent P and Clairon A 1998 Phys. Rev. A 57 436 [13] Krzysztof S, Sang E P, Giuseppe M and Witold C 2010 Metrologia 47 363 [14] Kumagai M, Ito H, Kajita M and Hosokawa M 2008 Metrologia 45 139 [15] Rosenbusch P, Zhang S and Clairon A 2007 IEEE International Frequency Control Symposium, 2007 Joint with the 21st European Frequency and Time Forum p 1060
[1]
. [J]. 中国物理快报, 2022, 39(8): 80601-.
[2]
. [J]. 中国物理快报, 2022, 39(4): 44201-.
[3]
. [J]. 中国物理快报, 2020, 37(9): 94201-.
[4]
. [J]. 中国物理快报, 2019, 36(12): 120601-.
[5]
. [J]. 中国物理快报, 2019, 36(7): 70601-.
[6]
. [J]. 中国物理快报, 2018, 35(8): 80601-.
[7]
. [J]. 中国物理快报, 2017, 34(10): 109901-.
[8]
. [J]. 中国物理快报, 2017, 34(9): 90601-.
[9]
. [J]. 中国物理快报, 2017, 34(9): 90602-.
[10]
. [J]. 中国物理快报, 2017, 34(5): 50601-.
[11]
. [J]. 中国物理快报, 2017, 34(2): 20601-020601.
[12]
. [J]. 中国物理快报, 2016, 33(11): 114202-114202.
[13]
. [J]. 中国物理快报, 2016, 33(07): 70601-070601.
[14]
. [J]. 中国物理快报, 2016, 33(07): 70602-070602.
[15]
. [J]. 中国物理快报, 2016, 33(04): 40601-040601.