Remeasurement of the Half-Life of 79Se with the Projectile X-Ray Detection Method

  • Remeasurement of the half-life of 79Se has been carried out with the projectile x-ray detection (PXD) in accelerator mass spectrometry. We made two major improvements over our earlier 79Se half-life determination Nucl. Instrum. Methods B 123 (1997) 403. Fristly, the PXD technique has been used for separation of 79Se and 75Se from 79Br and 75As isobars respectively. Secondly, the half-life of 79Se has been measured relative to the known precise half-life of 75Se (119.79 ± 0.04d). A more reliable half-life of 79Se measured with the PXD technique is (2.95 ± 0.38)x105y, about a factor of one lower than the previous value, 1.1 x 106y. The problems in the previous measurement are discussed.