Measurement of Electron-Drift Velocity in Ar+CH4 Mixtures Using Double-Grid Method

  • Based on analyzing the induced signals from the double-grids of an ionization chamber, the electron-drift time between the two grids is determined and the electron-drift velocity is derived. A waveform digitizer is employed to record pulses from the two grids of the ionization chamber. The electron-drift velocity is measured as a function of the reduced electric field E/p for eight different ratios of Ar+CH4 mixtures. By analyzing the experimental data of this study, self-consistency of experimental data is achieved, and formulae for calculating electron-drift velocity in any ratio of Ar+CH4 mixtures are obtained.
  • Article Text

  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return