Modified Method for Efficiency Calibration of High Energy γ Detector |
YE Zongyuan;LI Yubing;DING Shengyao;BAO Zongyu;YANG Xiaoyun;RONG Chaofan;DING Xixiang;ZHENG Jinmei |
Division of Radiometrology, China Institute of Atomic Energy, Beijing 102413 |
|
Cite this article: |
YE Zongyuan, LI Yubing, DING Shengyao et al 1994 Chin. Phys. Lett. 11 12-15 |
|
|
Abstract A modified method of absolute efficiency calibration for germanium detector has been developed based on the fact that the intensities of capture γ-rays can be determined by the 1368KeV γ-rays from sodium activation when a NaCI sample is irradiated by thermal neutrons. As a verification of this method, the intensity of the 4.44MeV γ-ray from an Am-Be neutron source was determined.
|