Abstract:Lesions in porcine liver tissues created by continuous high intensity focused ultrasound (HIFU) exposures in vitro are theoretically and experimentally investigated, with the transmitter moving along a linear path at a fixed speed. Numerical simulations of the lesion formation are performed based on the Khokhlov–Zabolotskaya–Kuznetov equation and the bio-heat equation. In order to verify the theoretical predictions, experiments are performed in the one-dimensional scanning mode to measure the cross-sectional area of lesions created in the in vitro porcine liver exposed to 1.01-MHz HIFU pulses with the acoustic power of 70 W. The results indicate that, compared to the traditional discrete treatment protocol, the application of a continuous scanning model can create more uniform lesions in tissues and significantly reduces the total treatment time from 47 s to 30 s.